High-burst-strength, feedback-controlled bipolar vessel sealing
- PMID: 9602010
- DOI: 10.1007/s004649900733
High-burst-strength, feedback-controlled bipolar vessel sealing
Abstract
Background: The inherent tedium of intracorporeal knot tying has stimulated greater interest in energy-based and mechanical alternatives for hemostasis.
Methods: Three hundred thirty-one arteries and veins were sealed by application of precisely controlled electrothermal energy and physical pressure, allowing for brief cooling in compression, in experimental animals and fresh abattoir vessels. These seals were compared for bursting strength with occlusions by ultrasonic and bipolar coagulation, surgical clips, and ligatures.
Results: Ultrasonic and bipolar occlusions were significantly less likely to have burst strengths greater than 400 mmHg as compared with seals, clips, and ligatures (p < 0.001). Seal competence could be visually assessed by its translucence.
Conclusions: Precise energy control with physical compression, including a brief cooldown, produces a distinctive, translucent seal of partially denatured protein that can typically be transected after a single application. These seals have bursting strengths comparable to those of clips and ligatures and resist dislodgement because they are intrinsic to the vessel wall structure.
Similar articles
-
Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries.Surg Endosc. 2003 Aug;17(8):1228-30. doi: 10.1007/s00464-002-8833-7. Epub 2003 Jun 13. Surg Endosc. 2003. PMID: 12799888
-
Effectiveness of the ultrasonic coagulating shears, LigaSure vessel sealer, and surgical clip application in biliary surgery: a comparative analysis.Am Surg. 2001 Sep;67(9):901-6. Am Surg. 2001. PMID: 11565773
-
Validation of a Laparoscopic Ferromagnetic Technology-based Vessel Sealing Device and Comparative Study to Ultrasonic and Bipolar Laparoscopic Devices.Surg Laparosc Endosc Percutan Tech. 2017 Apr;27(2):e12-e17. doi: 10.1097/SLE.0000000000000385. Surg Laparosc Endosc Percutan Tech. 2017. PMID: 28234706 Free PMC article.
-
Novel hemostatic devices in thyroid surgery: electrothermal bipolar vessel sealing system and harmonic scalpel.Expert Rev Med Devices. 2008 Jul;5(4):447-66. doi: 10.1586/17434440.5.4.447. Expert Rev Med Devices. 2008. PMID: 18573045 Review.
-
A review of currently available vessel sealing systems.Minim Invasive Ther Allied Technol. 2007;16(1):52-7. doi: 10.1080/13645700601181414. Minim Invasive Ther Allied Technol. 2007. PMID: 17365677 Review.
Cited by
-
Early Experience with LigaSure Thyroidectomy in a Nigeria Teaching Hospital.Niger J Surg. 2019 Jan-Jun;25(1):64-69. doi: 10.4103/njs.NJS_40_18. Niger J Surg. 2019. PMID: 31007515 Free PMC article.
-
Real-Time, Nondestructive Optical Feedback Systems for Infrared Laser Sealing of Blood Vessels.Proc SPIE Int Soc Opt Eng. 2022 Jan-Feb;11936:1193605. doi: 10.1117/12.2605600. Epub 2022 Mar 3. Proc SPIE Int Soc Opt Eng. 2022. PMID: 35949201 Free PMC article.
-
LigaSure in total thyroidectomy.Surg Today. 2008;38(6):495-8. doi: 10.1007/s00595-007-3665-7. Epub 2008 May 31. Surg Today. 2008. PMID: 18516527
-
Monitoring Acute Stroke in Mouse Model Using Laser Speckle Imaging-Guided Visible-Light Optical Coherence Tomography.IEEE Trans Biomed Eng. 2018 Oct;65(10):2136-2142. doi: 10.1109/TBME.2017.2706976. Epub 2017 May 23. IEEE Trans Biomed Eng. 2018. PMID: 28541195 Free PMC article.
-
[Bleeding complications in bariatric surgery: Prophylaxis and therapy].Chirurg. 2015 Sep;86(9):833-40. doi: 10.1007/s00104-015-0038-4. Chirurg. 2015. PMID: 26099290 Review. German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical