Interactive MR imaging-guided biopsy and aspiration with a modified clinical C-arm system
- PMID: 9609180
- DOI: 10.2214/ajr.170.6.9609180
Interactive MR imaging-guided biopsy and aspiration with a modified clinical C-arm system
Abstract
Objective: This investigation was performed to test the hypotheses that interactive guidance of MR image acquisition during needle-directed procedures using a clinical 0.2-T C-arm open MR imaging system integrated with a frameless optically linked stereotaxy system is feasible, and that procedure times can be sufficiently short to be well tolerated by the patient.
Subjects and methods: One hundred six MR-guided procedures were performed in 86 patients (ranging in age from 5 months to 88 years) using a clinical C-arm imaging system supplemented with an in-room RF-shielded liquid crystal display monitor, a frameless stereotaxy system, rapid gradient-echo sequences for needle guidance, and MR-compatible monitoring and surgical lighting equipment. We performed 50 biopsies and aspirations of the head and neck in 37 patients, 23 biopsies of musculoskeletal lesions in 22 patients, 16 biopsies of abdominal sites in 10 patients, six biopsies of the thoracolumbar spine or sacrum in six patients, and 11 shoulder joint injections for MR arthrography in 11 patients, in addition to 38 MR arthrographic injections on the same imaging system described in a previous report. Tissue sampling included fine-needle aspiration (n = 90) and cutting needle core biopsy (n = 41). Thirty-five patients underwent both procedures. Procedures were evaluated for success of needle placement, procedure time, and complications.
Results: Needle placement was successful in all cases, and no complications occurred. Tissue was sufficient for pathologic diagnosis for all but eight patients. Passes per patient averaged 2.1. For fine-needle aspiration, instrument time averaged 7 min 42 sec per pass, cutting needle core biopsy averaged 6 min 24 sec, and shoulder injection averaged 8 min.
Conclusion: MR imaging guidance for needle procedures on a clinical 0.2-T C-arm system with supplemental interventional accessories is feasible, with relatively rapid needle placement.
Similar articles
-
Percutaneous magnetic resonance image-guided biopsy and aspiration in the head and neck.Laryngoscope. 2000 Mar;110(3 Pt 1):382-5. doi: 10.1097/00005537-200003000-00009. Laryngoscope. 2000. PMID: 10718423
-
Interventional musculoskeletal procedures performed by using MR imaging guidance with a vertically open MR unit: assessment of techniques and applicability.Radiology. 2002 Apr;223(1):127-36. doi: 10.1148/radiol.2231010900. Radiology. 2002. PMID: 11930057
-
[Interactive stereotaxic interventions in superconducting, open 0.5-Tesla MRI tomography].Rofo. 1996 Sep;165(3):276-80. doi: 10.1055/s-2007-1015755. Rofo. 1996. PMID: 8924689 German.
-
Musculoskeletal interventional MR imaging.Magn Reson Imaging Clin N Am. 2005 Aug;13(3):519-32. doi: 10.1016/j.mric.2005.04.010. Magn Reson Imaging Clin N Am. 2005. PMID: 16084417 Review.
-
Percutaneous biopsy from blinded to MR guided: an update on current techniques and applications.Magn Reson Imaging Clin N Am. 2005 Aug;13(3):441-64. doi: 10.1016/j.mric.2005.04.009. Magn Reson Imaging Clin N Am. 2005. PMID: 16084412 Review.
Cited by
-
Real-time MR fluoroscopy-navigated lumbar facet joint injections: feasibility and technical properties.Eur Radiol. 2008 Jul;18(7):1513-8. doi: 10.1007/s00330-008-0890-4. Epub 2008 Feb 26. Eur Radiol. 2008. PMID: 18299836
-
Perk Station--Percutaneous surgery training and performance measurement platform.Comput Med Imaging Graph. 2010 Jan;34(1):19-32. doi: 10.1016/j.compmedimag.2009.05.001. Epub 2009 Jun 17. Comput Med Imaging Graph. 2010. PMID: 19539446 Free PMC article.
-
Glenohumeral joint injection: a comparative study of ultrasound and fluoroscopically guided techniques before MR arthrography.Eur Radiol. 2009 Mar;19(3):722-30. doi: 10.1007/s00330-008-1200-x. Epub 2008 Oct 29. Eur Radiol. 2009. PMID: 18958474 Free PMC article.
-
MRI-guided abdominal biopsy in a 0.23-T open-configuration MRI system.Eur Radiol. 2005 Jun;15(6):1256-62. doi: 10.1007/s00330-004-2566-z. Epub 2004 Dec 31. Eur Radiol. 2005. PMID: 15627187
-
Feasibility of MR-temperature mapping of ultrasonic heating from a CMUT.IEEE Trans Ultrason Ferroelectr Freq Control. 2008 Apr;55(4):811-8. doi: 10.1109/TUFFC.2008.715. IEEE Trans Ultrason Ferroelectr Freq Control. 2008. PMID: 18467225 Free PMC article.
Publication types