Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 1998 Jun;109(6):748-53.
doi: 10.1093/ajcp/109.6.748.

Evaluation of a new rapid D-dimer assay for clinically suspected deep venous thrombosis (Liatest D-dimer)

Affiliations
Comparative Study

Evaluation of a new rapid D-dimer assay for clinically suspected deep venous thrombosis (Liatest D-dimer)

M Escoffre-Barbe et al. Am J Clin Pathol. 1998 Jun.

Abstract

In previous studies, enzyme-linked immunosorbent assays (ELISA) for plasma D-dimer analysis have demonstrated high sensitivity, suggesting their potential usefulness in excluding deep venous thrombosis (DVT). We evaluated the usefulness of a new D-dimer test (Liatest D-dimer) for suspected DVT in a prospective study of patients admitted to the hospital because of recent (not exceeding 1 week before admission) clinical signs. Contrast venography or compression ultrasonography or both were performed within 24 hours of admission. A new quantitative determination of D-dimer concentration using a suspension of microlatex particles coated with specific antibodies was tested. A standard plasma D-dimer ELISA measurement was also performed. Of 464 patients, 276 had a proven DVT (distal, 74; proximal, 202). For a cutoff level of 400 ng/mL, sensitivity of the Liatest method in the diagnosis of overall DVT was 94.6% (95% confidence interval, 92.0%-97.0%), and the specificity was 35% (95% confidence interval, 28%-42%). The sensitivity and negative predictive value were 98.5% and 95.6%, respectively, in the diagnosis of proximal DVT, but only 83.8% and 84.6%, respectively, in the diagnosis of distal DVT. This new rapid Liatest D-dimer assay seems to be highly sensitive and could replace the ELISA method in excluding patients with proximal DVT. Both methods provide lower sensitivity for distal DVT.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources