Hemoglobin Raleigh as the cause of a falsely increased hemoglobin A1C in an automated ion-exchange HPLC method
- PMID: 9625056
Hemoglobin Raleigh as the cause of a falsely increased hemoglobin A1C in an automated ion-exchange HPLC method
Abstract
Irreversible glycation of the hemoglobin A0 (HbA0) beta chain leads to the production of HbA1C, which can be used to monitor long-term blood glucose control in patients with diabetes mellitus. HbA1C is less positively charged than nonglycated HbA0, and this decrease in charge is the basis of ion-exchange and electrophoretic methods that measure HbA1C. We recently identified a sample that appeared to contain 46% HbA1C by an automated ion-exchange HPLC method (Bio-Rad Variant) but only 3.8% by an immunoinhibition latex agglutination method. A combination of traditional and mass spectrometric protein analysis and genomic DNA analysis of the Hb beta chain and genes revealed that the patient was heterozygotic for Hb-Raleigh, a variant containing a valine-->alanine substitution at position 1 of the beta chain. The amino-terminal alanine in this variant Hb is posttranslationally modified by acetylation, leading to a charge difference similar to glycation and making the behavior of HbA1C and Hb Raleigh virtually identical in the ion-exchange HPLC method. This observation suggests that it is important to confirm HbA1C values in excess of 15%, especially if they are not consistent with the clinical picture, by an independent HbA1C method such as immunoassay or boronic acid affinity chromatography. However, for this particular variant Hb, even these latter methods might be misleading, because the acetylated N-terminal amino acid of the Hb-Raleigh beta chain cannot be glycated.
Similar articles
-
Hemoglobin Raleigh results in factitiously low hemoglobin A1c when evaluated via immunoassay analyzer.J Diabetes Complications. 2011 Jan-Feb;25(1):14-8. doi: 10.1016/j.jdiacomp.2009.09.004. Epub 2009 Nov 6. J Diabetes Complications. 2011. PMID: 19896869
-
Mass spectrometry based characterization of Hb Beckman variant in a falsely elevated HbA(1c) sample.Anal Biochem. 2015 Nov 15;489:53-8. doi: 10.1016/j.ab.2015.07.010. Epub 2015 Jul 23. Anal Biochem. 2015. PMID: 26209877
-
Sky High or Undetectable? A Patient with Discordant Hemoglobin A1c.Lab Med. 2021 Sep 1;52(5):e129-e132. doi: 10.1093/labmed/lmab012. Lab Med. 2021. PMID: 33724425
-
Effects of hemoglobin variants and chemically modified derivatives on assays for glycohemoglobin.Clin Chem. 2001 Feb;47(2):153-63. Clin Chem. 2001. PMID: 11159762 Review.
-
Pathology consultation on HbA(1c) methods and interferences.Am J Clin Pathol. 2014 Jan;141(1):5-16. doi: 10.1309/AJCPQ23GTTMLAEVL. Am J Clin Pathol. 2014. PMID: 24343732 Review.
Cited by
-
Usefulness of glycated albumin assay for diabetes monitoring.J Diabetes Sci Technol. 2011 Nov 1;5(6):1463-5. doi: 10.1177/193229681100500620. J Diabetes Sci Technol. 2011. PMID: 22226266 Free PMC article.
-
Characterization of the elusive disulfide bridge forming human Hb variant: Hb Ta-Li beta83 (EF7)Gly --> Cys by electrospray mass spectrometry.J Am Soc Mass Spectrom. 2002 Feb;13(2):187-91. doi: 10.1016/s1044-0305(01)00349-x. J Am Soc Mass Spectrom. 2002. PMID: 11838022
-
Impact of haemoglobin variants on the use of haemoglobin A1c for the diagnosis and monitoring of diabetes: a contextualised review.Ir J Med Sci. 2023 Feb;192(1):169-176. doi: 10.1007/s11845-022-02967-2. Epub 2022 Apr 1. Ir J Med Sci. 2023. PMID: 35362846 Free PMC article. Review.
-
Interpretation of HbA1c lies at the intersection of analytical methodology, clinical biochemistry and hematology (Review).Exp Ther Med. 2022 Oct 4;24(6):707. doi: 10.3892/etm.2022.11643. eCollection 2022 Dec. Exp Ther Med. 2022. PMID: 36382101 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical