Accuracy of two pulse oximeters at low arterial hemoglobin-oxygen saturation
- PMID: 9635666
- DOI: 10.1097/00003246-199806000-00040
Accuracy of two pulse oximeters at low arterial hemoglobin-oxygen saturation
Abstract
Objective: To evaluate the performance of two pulse oximeters in the measurement of arterial hemoglobin saturation in hypoxemic children.
Design: Prospective, repeated-measures observational study.
Setting: A 16-bed pediatric intensive care unit in a children's tertiary hospital.
Patients: Sixty-six patients with arterial saturation of <90%.
Interventions: Three arterial blood samples were taken from each subject during a 48-hr period. Pulse oximeter measurements of arterial saturation were compared with arterial saturation determined by cooximetry.
Measurements and main results: Arterial saturation was measured using one or both pulse oximeters (SpO2) and compared with the arterial hemoglobin saturation determined by cooximetry (SaO2). Sixty-two subjects were studied, using the Ohmeda pulse oximeter giving 185 data points (78 with saturations <75% [defined by the average of pulse oximeter and cooximeter]); 53 subjects were studied, using the Hewlett-Packard pulse oximeter yielding 155 data points (60 with saturations <75%). SpO2 ranged from 24% to 94%. Bias and precision of the Ohmeda pulse oximeter were -2.8% and 4.8% >75% and -0.8% and 8.0% <75%. Bias and precision of the Hewlett-Packard pulse oximeter were -0.5% and 5.1% >75% and 0.4% and 4.6% <75%. Intrapatient regression coefficient (r) for the differences between pulse oximeter and cooximeter was 0.58 for the Ohmeda and 0.59 for the Hewlett-Packard. Regression coefficients for predicting change in cooximeter value given a change in the Ohmeda pulse oximeter were 0.59 and 0.71 <75% and >75%, respectively. Similar coefficients for the Hewlett-Packard pulse oximeter were 0.50 and 0.70, respectively.
Conclusion: The performance of the Ohmeda pulse oximeter deteriorated below an SpO2 of 75%. The Hewlett-Packard pulse oximeter performed consistently above and below an SpO2 of 75%. The ability of both pulse oximeters to reliably predict change in SaO2 based on change in pulse oximetry was limited. We recommend measurement of PaO2 or SaO2 for important clinical decisions.
Similar articles
-
Saturation by pulse oximetry: comparison of the results obtained by instruments of different brands.J Pediatr. 1993 Apr;122(4):620-6. doi: 10.1016/s0022-3476(05)83549-8. J Pediatr. 1993. PMID: 7681875
-
Simultaneous comparison of three noninvasive oximeters in healthy volunteers.Med Instrum. 1987 Jun;21(3):183-8. Med Instrum. 1987. PMID: 3614041
-
A comparison of a new reflectance oximeter with the Hewlett-Packard ear oximeter.Biomed Instrum Technol. 1990 Mar-Apr;24(2):122-6. Biomed Instrum Technol. 1990. PMID: 2322755
-
Accuracy of laminated disposable pulse-oximeter sensors.Respir Care. 1995 Jul;40(7):728-33. Respir Care. 1995. PMID: 10144067 Review.
-
Impacts of Skin Color and Hypoxemia on Noninvasive Assessment of Peripheral Blood Oxygen Saturation: A Scoping Review.Cureus. 2023 Sep 27;15(9):e46078. doi: 10.7759/cureus.46078. eCollection 2023 Sep. Cureus. 2023. PMID: 37900526 Free PMC article.
Cited by
-
Racial and ethnical discrepancy in hypoxemia detection in patients on extracorporeal membrane oxygenation.JTCVS Open. 2023 Mar 1;14:145-170. doi: 10.1016/j.xjon.2023.02.011. eCollection 2023 Jun. JTCVS Open. 2023. PMID: 37425474 Free PMC article.
-
Assessment of a Non Invasive Brain Oximeter in Volunteers Undergoing Acute Hypoxia.Med Devices (Auckl). 2020 Jun 30;13:183-194. doi: 10.2147/MDER.S250102. eCollection 2020. Med Devices (Auckl). 2020. PMID: 32669881 Free PMC article.
-
Comparative Analysis of Oxygen Saturation by Pulse Oximetry and Arterial Blood Gas in Hypoxemic Patients in a Tertiary Care Hospital.Cureus. 2023 Jul 25;15(7):e42447. doi: 10.7759/cureus.42447. eCollection 2023 Jul. Cureus. 2023. PMID: 37637606 Free PMC article.
-
Characterizing the Racial Discrepancy in Hypoxemia Detection in Venovenous Extracorporeal Membrane Oxygenation: An Extracorporeal Life Support Organization Registry Analysis.Lung. 2024 Aug;202(4):471-481. doi: 10.1007/s00408-024-00711-4. Epub 2024 Jun 10. Lung. 2024. PMID: 38856932 Free PMC article.
-
Assessment of a Non-Invasive Brain Oximeter in a Sheep Model of Acute Brain Injury.Med Devices (Auckl). 2019 Dec 3;12:479-487. doi: 10.2147/MDER.S235804. eCollection 2019. Med Devices (Auckl). 2019. PMID: 31824197 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources