Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 1998 Jul;70(1):152-4.
doi: 10.1016/s0015-0282(98)00101-0.

Reliability of intraobserver and interobserver sonographic endometrial stripe thickness measurements

Affiliations
Clinical Trial

Reliability of intraobserver and interobserver sonographic endometrial stripe thickness measurements

S D Spandorfer et al. Fertil Steril. 1998 Jul.

Abstract

Objective: To determine the intraobserver and interobserver reliability of endometrial stripe thickness measurements in women undergoing controlled ovarian hyperstimulation.

Design: Prospective blinded study.

Setting: Tertiary care, university hospital.

Patient(s): Sixty-three patients undergoing controlled ovarian hyperstimulation and being monitored with transvaginal ultrasound were studied.

Intervention(s): None.

Main outcome measure(s): Intraobserver and interobserver variability of endometrial stripe thickness measurements between 2 blinded observers, with 2 observations made by each observer.

Result(s): A statistically significant correlation was detected between the 2 measurements of each observer. The mean (+/-SD) interobserver difference between the average of the 2 measurements performed by both observers was 1.02 +/- 0.82 mm. A statistically significant correlation between the measurements of the 2 observers was detected. For both observers A and B, who used < or =6 mm as an abnormal endometrial thickness, an excellent level of agreement was detected between the 2 measurements made on each patient. When comparing the average values obtained by the 2 observers for each of the patients, an excellent level of agreement was detected.

Conclusion(s): There is an excellent correlation between intraobserver and interobserver measurements of endometrial stripe thickness.

PubMed Disclaimer

Publication types