Comparative Study
. 1998 Jun;45(6):578-83.
doi: 10.1007/BF03012713.

Evaluation of a new semi-continuous cardiac output system in the intensive care unit

Affiliations
Comparative Study

Evaluation of a new semi-continuous cardiac output system in the intensive care unit

P Seguin et al. Can J Anaesth. 1998 Jun.

Abstract

Purpose: A new semi-continuous thermodilution cardiac output (CCO) system has been developed recently (Opti-Q and Q-vue Abbott critical care system). The aim of this study was to compare the accuracy and reproducibility of this new device with conventional ice-bolus thermodilution cardiac output (BCO).

Methods: Fifteen critically ill patients who needed pulmonary artery catheterization were prospectively investigated. Eighty seven paired data using BCO and CCO methods were compared. Reproducibility was assessed from 90 BCO and 87 CCO determinations by calculation of the mean standard error (SEM) and according to Bland and Altman methodology.

Results: The BCO and CCO ranged from 2.46 to 11.20 L.min-1 and from 1.75 to 10.05 L.min-1 respectively. Bias (mean difference between BCO and CCO) was null (0.002 L.min-1, P = 0.98), precision (SD of the bias) was 0.74 L.min-1 and the limits of agreement (mean difference +/- 1.96 SD) ranged from -1.45 to 1.45 L.min-1. The threshold to consider two cardiac outputs as different (3 x SEM) was equivalent for BCO and CCO (0.54 and 0.465 L.min-1 respectively). According to the Bland and Altman method, reproducibility of CCO was greater than that of BCO; bias of repeated measurements of BCO and CCO were 0.15 L.min-1 (P < 0.05) and 0.047 L.min-1 (NS), respectively.

Conclusion: Compared with BCO, this new device was accurate but cannot be considered as interchangeable regarding the limits of agreement. Reproducibility of CCO was superior to BCO.