Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 1998 Jul;25(7 Pt 1):1111-7.
doi: 10.1118/1.598433.

Scattered radiation in scanning slot mammography

Affiliations
Comparative Study

Scattered radiation in scanning slot mammography

Z Jing et al. Med Phys. 1998 Jul.

Abstract

Monte Carlo simulations were used to quantify the amount of scattered radiation a scanning slot detector geometry designed for use in digital mammography. Ratios of the scatter to primary (S/P) x-ray photon energy absorbed in the detector were obtained for a Lucite phantom, and were investigated as a function of photon energy, phantom thickness, and slot detector width. Over a Lucite phantom thickness range of 2-6 cm, the S/P ratios range from about 0.10 to 0.17 for a 4 mm wide slot detector at the x-ray photon energies used in mammography. These ratios increased by a factor of approximately 1.8 when the slot width was increased to 10 mm. In general, 20 keV photons gave S/P ratios similar to those of a 30 kVp x-ray spectrum (Mo target + 30 microns Mo filtration). The use of a 3 cm air gap reduced the S/P ratios by a factor of between 2.5 and 3.4, depending on the phantom thickness. For a constant primary energy fluence, coherent scatter was reduced as photon energy increased, whereas Compton scatter increased with increasing photon energy. With no air gap, the contributions of coherent and Compton scatter were found to be equal at 25 keV, whereas the introduction of a 3 cm air gap resulted in equal contributions for the two scatter processes at 36 keV. A 10 mm wide slot detector consisting of a 36.7 mg/cm2 thick Gd2O2S:Tb phosphor screen was compared to an ideal detector absorbing all incident primary/scatter photons. Average differences in the S/P ratios for these two detectors were 7% with no air gap and approximately 4% with a 3 cm air gap. The results obtained in this study will assist in the design of an optimal slot detector for use in digital mammography.

PubMed Disclaimer

Publication types

LinkOut - more resources