Effects of physiologic motion of the human brain upon quantitative 1H-MRS: analysis and correction by retro-gating
- PMID: 9699493
- DOI: 10.1002/(sici)1099-1492(199805)11:3<107::aid-nbm525>3.0.co;2-i
Effects of physiologic motion of the human brain upon quantitative 1H-MRS: analysis and correction by retro-gating
Abstract
Signal loss and absolute quantitation errors in 1H-MRS (localized proton MR spectroscopy) because of physiologic brain motion are analyzed quantitatively. Cardiac and respiratory related motion lead to substantial phase dispersion when using a standard, short echo-time STEAM sequence. The loss in signal area varies from 6-7% with TM (middle interval time in a STEAM sequence) = 13.7 ms, to 25-39% with TM = 100 ms. The variation in signal area because of motion-related phase dispersion is up to 16% for TM = 100 ms. The signal phase as a function of the position in the cardiac cycle is shown to be reproducible. Maximal differences in the signal phase are over 180 degrees for long TMs. ECG-gating reduces the phase dispersion considerably but introduces problems with variable repetition times. Using a phase calibration curve recorded with the water suppression turned off, it is possible to retrogate subsequent untriggered acquisitions with the water suppression activated, if the time points in the cardiac cycle are recorded for each acquisition. The gain in signal intensity is between 3 and 21%. For absolute quantification via brain water, this phase analysis has the important consequence that reference scans must be phased individually before co-adding, otherwise metabolite concentrations may be severely overestimated.
Similar articles
-
Quantitative analysis of the effects of physiologic brain motion on point-resolved spectroscopy.AJNR Am J Neuroradiol. 2006 May;27(5):1070-3. AJNR Am J Neuroradiol. 2006. PMID: 16687544 Free PMC article.
-
Restoration of motion-related signal loss and line-shape deterioration of proton MR spectra using the residual water as intrinsic reference.Magn Reson Med. 2001 Aug;46(2):395-400. doi: 10.1002/mrm.1203. Magn Reson Med. 2001. PMID: 11477645
-
Metabolite-cycled STEAM and semi-LASER localization for MR spectroscopy of the human brain at 9.4T.Magn Reson Med. 2018 Apr;79(4):1841-1850. doi: 10.1002/mrm.26873. Epub 2017 Aug 15. Magn Reson Med. 2018. PMID: 28812315
-
1H MR spectroscopy of the brain: absolute quantification of metabolites.Radiology. 2006 Aug;240(2):318-32. doi: 10.1148/radiol.2402050314. Radiology. 2006. PMID: 16864664 Review.
-
In vivo quantitation of metabolite concentrations in the brain by means of proton MRS.NMR Biomed. 1995 Jun;8(4):139-48. doi: 10.1002/nbm.1940080402. NMR Biomed. 1995. PMID: 8771088 Review.
Cited by
-
A novel phase and frequency navigator for proton magnetic resonance spectroscopy using water-suppression cycling.Magn Reson Med. 2011 Jan;65(1):13-7. doi: 10.1002/mrm.22582. Magn Reson Med. 2011. PMID: 20872862 Free PMC article.
-
Fast and motion-robust saturation transfer MRI with inherent B0 correction using rosette trajectories and compressed sensing.Magn Reson Med. 2024 Dec;92(6):2535-2545. doi: 10.1002/mrm.30249. Epub 2024 Aug 11. Magn Reson Med. 2024. PMID: 39129199
-
Effects of physiologic human brain motion on proton spectroscopy: quantitative analysis and correction with cardiac gating.AJNR Am J Neuroradiol. 2002 Feb;23(2):225-30. AJNR Am J Neuroradiol. 2002. PMID: 11847046 Free PMC article.
-
Motion correction in magnetic resonance spectroscopy.Magn Reson Med. 2020 Nov;84(5):2312-2326. doi: 10.1002/mrm.28287. Epub 2020 Apr 17. Magn Reson Med. 2020. PMID: 32301174 Free PMC article. Review.
-
Advanced MRI in multiple sclerosis: current status and future challenges.Neurol Clin. 2011 May;29(2):357-80. doi: 10.1016/j.ncl.2010.12.011. Neurol Clin. 2011. PMID: 21439446 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources