Accuracy of blood pressure measurement in the family practice center
- PMID: 9719346
- DOI: 10.3122/jabfm.11.4.252
Accuracy of blood pressure measurement in the family practice center
Abstract
Background: Equipment, physiologic, and technique factors can influence the accuracy of blood pressure measurement. The current study was designed to isolate certain technique factors and then assess the accuracy of nursing personnel measurements of blood pressure in three residency family practice centers.
Methods: An experienced registered nurse was trained in applying the American Heart Association recommendations for determining blood pressure by sphygmomanometry; three 1.5-hour practice sessions demonstrated her accuracy. Nine full days were then spent in the family practice centers rechecking as many staff blood pressure readings as possible while controlling for confounding variables.
Results: The following findings were significant: (1) the average absolute differences between control and study nurse systolic and diastolic blood pressure readings were 6.2 mmHg and 4.7 mmHg, respectively; (2) a unidirectional error of 3.8 mmHg in the measurement of diastolic blood pressure was found in one center, possibly because less care was taken with afternoon measurements; (3) variability in systolic blood pressure readings was higher in all three centers (+/- 8.5 mmHg) than attained during the training sessions for the control nurse (+/- 5.8 mmHg); (4) the average errors attributable to technique factors studied that were potentially correctable by training were only 1.8 mmHg for systolic and 0.7 mmHg for diastolic pressures.
Conclusions: The degree of inaccuracy in ambulatory nursing blood pressure readings attributable to errors in technique is quantified by this study. Training can reduce, but not abolish, this inaccuracy. Careful attention to proper blood pressure measurement technique and such variables as equipment calibration is essential for both nursing and physician observers. Taking multiple blood pressure measurements before making clinical decisions can limit the effect of these inaccuracies.
Similar articles
-
Self-measurement of blood pressure: accuracy, patient preparation for readings, technique and equipment.Blood Press Monit. 2001 Jun;6(3):133-8. doi: 10.1097/00126097-200106000-00003. Blood Press Monit. 2001. PMID: 11518835
-
The trained observer better predicts daytime ABPM diastolic blood pressure in hypertensive patients than does an automated (Omron) device.Blood Press Monit. 2006 Apr;11(2):53-8. doi: 10.1097/01.mbp.0000200480.26669.72. Blood Press Monit. 2006. PMID: 16534405
-
[Quality assurance in Riva-Rocci blood pressure measurement: simultaneous sphygmomanometry with open and covered pressure display].Z Kardiol. 1988 Aug;77(8):537-42. Z Kardiol. 1988. PMID: 3176598 German.
-
A review of common errors in the indirect measurement of blood pressure. Sphygmomanometry.Arch Intern Med. 1993 Dec 27;153(24):2741-8. Arch Intern Med. 1993. PMID: 8257249 Review.
-
Development of diagnostic thresholds for automated self-measurement of blood pressure in adults. First International Consensus Conference on Blood Pressure Self-Measurement.Blood Press Monit. 2000 Apr;5(2):101-9. Blood Press Monit. 2000. PMID: 10828897 Review.
Cited by
-
Comparing the effects of white coat hypertension and sustained hypertension on mortality in a UK primary care setting.Ann Fam Med. 2008 Sep-Oct;6(5):390-6. doi: 10.1370/afm.865. Ann Fam Med. 2008. PMID: 18779542 Free PMC article.
-
Nurse effects on measurement error in household biosocial surveys.BMC Med Res Methodol. 2020 Feb 27;20(1):45. doi: 10.1186/s12874-020-00922-2. BMC Med Res Methodol. 2020. PMID: 32106825 Free PMC article.
-
Validity of measures is no simple matter.Health Serv Res. 2005 Oct;40(5 Pt 2):1584-604. doi: 10.1111/j.1475-6773.2005.00443.x. Health Serv Res. 2005. PMID: 16178997 Free PMC article.
-
Screening for hypertension annually compared with current practice.Ann Fam Med. 2013 Mar-Apr;11(2):116-21. doi: 10.1370/afm.1467. Ann Fam Med. 2013. PMID: 23508597 Free PMC article.
-
A multi-level system quality improvement intervention to reduce racial disparities in hypertension care and control: study protocol.Implement Sci. 2013 Jun 4;8:60. doi: 10.1186/1748-5908-8-60. Implement Sci. 2013. PMID: 23734703 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources