Accuracy of reagent strips in detecting hypoglycemia in the emergency department
- PMID: 9737491
- DOI: 10.1016/s0196-0644(98)70005-8
Accuracy of reagent strips in detecting hypoglycemia in the emergency department
Abstract
Study objective: Although reagent strips are commonly used, their reliability to estimate blood glucose concentration and guide administration of dextrose solutions in the emergency department environment has not been proved. We determined the accuracy of visually interpreted reagent strips (Chemstrip bG, Boehringer Mannheim Corp, Indianapolis, IN) and their ability to identify hypoglycemic patients in the ED.
Methods: We conducted a prospective, nonrandomized blinded clinical study of the visual estimation of blood glucose values by ED personnel using Chemstrip bG reagent strips during a 4-month period. Simultaneously obtained blood samples sent for laboratory glucose determination served as controls. The study was conducted at a large university hospital ED with an urban patient population. A convenience sample of 215 adult ED patients underwent serum glucose determination with data form completion. No study intervention was tested, although timing of administration of dextrose solutions, if given, was recorded.
Results: Hypoglycemia was defined as a glucose concentration less than 60 mg/dL on standard laboratory analysis. Reagent strips identified 28 of 29 of these patients (sensitivity=97%), and 171 of 182 patients without hypoglycemia (specificity=94%, negative predictive value=99%) compared with control samples. The 1 false-negative reagent strip reading of 80 mg/dL was obtained from blood stored in a serum separator tube and had a laboratory glucose value of 39 mg/dL. Eighty-seven percent of the reagent strips were within +/-60 mg/dL of the control value for the laboratory glucose reference range less than 350 mg/dL.
Conclusion: Visually interpreted Chemstrip bG reagent strips provide an acceptable estimation of blood glucose concentration in the ED and are highly sensitive in detecting hypoglycemia.
Similar articles
-
A clinical comparison of the performance of four blood glucose reagent strips.Am J Emerg Med. 1990 Jan;8(1):11-5. doi: 10.1016/0735-6757(90)90286-9. Am J Emerg Med. 1990. PMID: 2293826
-
Comparison of Visidex and Chemstrip bG with Beckman Glucose Analyzer determination of blood glucose.Diabetes Care. 1983 Nov-Dec;6(6):536-9. doi: 10.2337/diacare.6.6.536. Diabetes Care. 1983. PMID: 6653310
-
Determination of prehospital blood glucose: a prospective, controlled study.J Emerg Med. 1992 Nov-Dec;10(6):679-82. doi: 10.1016/0736-4679(92)90524-w. J Emerg Med. 1992. PMID: 1491148
-
Glucose.Emerg Med Clin North Am. 1986 Feb;4(1):193-205. Emerg Med Clin North Am. 1986. PMID: 3512245 Review.
-
Glucose analysis: indications for ordering and alternatives to the laboratory.Ann Emerg Med. 1986 Mar;15(3):372-6. doi: 10.1016/s0196-0644(86)80588-1. Ann Emerg Med. 1986. PMID: 3511802 Review.