Comparative Study
. 1998 Oct;36(10):2853-60.
doi: 10.1128/JCM.36.10.2853-2860.1998.

Clinical evaluation of the automated COBAS AMPLICOR MTB assay for testing respiratory and nonrespiratory specimens

Affiliations
Comparative Study

Clinical evaluation of the automated COBAS AMPLICOR MTB assay for testing respiratory and nonrespiratory specimens

U Reischl et al. J Clin Microbiol. 1998 Oct.

Abstract

We evaluated the COBAS AMPLICOR PCR system (Roche Diagnostics) for the routine detection of Mycobacterium tuberculosis complex (MTBC) in clinical specimens. Diagnostic culture, considered as the reference method, was performed with BACTEC, Löwenstein-Jensen, Stonebrink, and Kirchner media. Occasionally MB-Redox, ESP, or MGIT medium was also used. A total of 643 respiratory and 506 nonrespiratory specimens collected from 807 patients were investigated. Of the 95 culture-positive specimens, 80 were COBAS AMPLICOR MTB positive, and of the 1,054 culture-negative specimens, 1,044 were COBAS AMPLICOR MTB negative. After resolving discrepancies by review of the medical history, the overall sensitivity, specificity, and positive and negative predictive values for the COBAS AMPLICOR MTB assay, respectively, were 83.5, 98.8, 86.7, and 98.6% compared to those of diagnostic culture. In smear-positive specimens, the sensitivity of the COBAS AMPLICOR MTB assay was 96%, versus 48% for smear-negative specimens. No significant differences in the test performance between respiratory and nonrespiratory specimens were observed. The overall inhibition rate was less than 2%, excluding stool specimens. The clear advantages of the COBAS AMPLICOR PCR system are standardized procedures and reagents for specimen processing as well as an internal control for reliable monitoring of PCR inhibitors. By simplifying the work flow through a completely automated amplification and amplicon detection procedure, the COBAS AMPLICOR PCR system proved itself as a very useful component for routine diagnostic procedures.

PubMed Disclaimer

Figures

FIG. 1
FIG. 1
Comparison of the workload organization of the COBAS AMPLICOR MTB assay (A) and a conventional PCR assay with that of an enzyme-linked oligosorbent assay-based amplicon detection (B). S, sample preparation; AMP, amplification; DP, detection process; MTB, analytic run.

Similar articles

Cited by

References

    1. Ausina V, Gamboa F, Gazapo E, Manterola J M, Lonca J, Matas L, Manzano J R, Rodrigo C, Cardona P J, Padilla E. Evaluation of the semiautomated Abbott LCx Mycobacterium tuberculosis assay for direct detection of Mycobacterium tuberculosis in respiratory specimens. J Clin Microbiol. 1997;35:1996–2002. - PMC - PubMed
    1. Beige J, Lokies J, Schaberg T, Finckh U, Fischer M, Mauch H, Lode H, Köhler B, Rolfs A. Clinical evaluation of a Mycobacterium tuberculosis PCR assay. J Clin Microbiol. 1995;33:90–95. - PMC - PubMed
    1. Böddinghaus B, Rogall T, Flohr T, Blöcker H, Böttger E C. Detection and identification of mycobacteria by amplification of rRNA. J Clin Microbiol. 1990;28:1751–1759. - PMC - PubMed
    1. Bodmer T, Gurtner A, Scholkmann M, Matter L. Evaluation of the COBAS AMPLICOR MTB system. J Clin Microbiol. 1997;35:1604–1605. - PMC - PubMed
    1. Carpentier E, Drouillard B, Dailloux M, Moinard D, Vallee E, Dutilh B, Maugein J, Bergogne-Berezin E, Carbonnelle B. Diagnosis of tuberculosis by Amplicor Mycobacterium tuberculosis test: a multicenter study. J Clin Microbiol. 1995;33:3106–3110. - PMC - PubMed

Publication types

MeSH terms

LinkOut - more resources