Comparative Study
. 1998 Aug;19(8):727-33.
doi: 10.1097/00006231-199808000-00002.

Volumetric measurements of bone mineral density of the lumbar spine: comparison of three geometrical approximations using dual-energy X-ray absorptiometry (DXA)

Affiliations
Comparative Study

Volumetric measurements of bone mineral density of the lumbar spine: comparison of three geometrical approximations using dual-energy X-ray absorptiometry (DXA)

M F Schreuder et al. Nucl Med Commun. 1998 Aug.

Abstract

Measurements of bone mineral density using dual-energy X-ray absorptiometry (DXA) gives area values (g cm-2) rather than true volumetric values (g cm-3). To calculate the vertebral volume using planar postero-anterior and lateral DXA values, several different geometrical approximations were used: cubic, cylindrical with a circular cross-section and cylindrical with an elliptical cross-section. The aim of this study was to compare these geometrical approximations with each other and with a reference standard, defined as the volume found on a computed tomographic (CT) scan. L2 and L3 were evaluated in a phantom study. Volume approximations by the cube or cylinder with circular cross-section geometry showed more than a 50% overestimation (range 54-74%). However, the elliptical cylinder approach showed very good agreement: 2.1% and 1.2% for L2 and L3, respectively, when compared to the CT volumes. In addition, we performed four patient studies with both CT and DXA to evaluate the elliptical cylinder estimate in a clinical setting. For L2 and L3, the mean relative difference was less than 2%. We conclude that the elliptical cylinder approach results in the most accurate bone volume estimates in both the phantom and patients.

PubMed Disclaimer

Comment in

Publication types

LinkOut - more resources