Use of a neonatal blood pressure cuff to monitor blood pressure in the adult finger--comparison with a standard adult arm cuff
- PMID: 9754611
- DOI: 10.1023/a:1009934032524
Use of a neonatal blood pressure cuff to monitor blood pressure in the adult finger--comparison with a standard adult arm cuff
Abstract
Background: There are few suitable methods for monitoring blood pressure continuously (or intermittently) for research in adult stroke patients, who are ill but do not justify invasive intensive care monitoring.
Method: We tested a neonatal arm blood pressure in adults by placing it on the forefinger ("finger cuff"). We compared the repeatability of the finger cuff with blood pressure measured by a standard adult arm cuff using the oscillometric technique in 168 ambulatory outpatients attending a cerebrovascular disease clinic.
Results: The mean difference between sequential mean blood pressure readings with the finger cuff was 0.55 mm Hg (95% confidence interval (CI) -14.36 to 15.47 mm Hg), and for the arm cuff was 3.31 mm Hg (95% CI -23.33 to 16.71 mm Hg). Measurements made with the arm cuff were shown to affect subsequent arm cuff readings made within a few minutes of the first. The mean difference between the finger cuff and arm cuff mean blood pressure readings was 0.03 mm Hg (95% CI -26.07 to 26.14 mm Hg) and agreement was better when the blood pressure was measured with the finger cuff first rather than the arm cuff. However, although there was no difference in the mean blood pressure recordings both systolic and diastolic blood pressure measurements differed systematically between arm and finger cuff.
Conclusion: The reproducibility of sequential blood pressure measurements made with the finger cuff was better than with the arm cuff. The performance of the finger cuff compared with that of the arm cuff was sufficiently good to encourage use of the finger cuff in research involving automatic intermittent monitoring to observe sequential blood pressures over time in stroke patients. However, measurements of systolic and diastolic pressure were not the same with the two cuffs and further work on calibration of the finger cuff would be useful.
Similar articles
-
The CNAP™ Finger Cuff for Noninvasive Beat-To-Beat Monitoring of Arterial Blood Pressure: An Evaluation in Intensive Care Unit Patients and a Comparison with 2 Intermittent Devices.Anesth Analg. 2016 Nov;123(5):1126-1135. doi: 10.1213/ANE.0000000000001324. Anesth Analg. 2016. PMID: 27159068
-
Can an electronic device with a single cuff be accurate in a wide range of arm size? Validation of the Visomat Comfort 20/40 device for home blood pressure monitoring.J Hum Hypertens. 2008 Nov;22(11):796-800. doi: 10.1038/jhh.2008.70. Epub 2008 Jul 3. J Hum Hypertens. 2008. PMID: 18596720
-
Blood pressure measurement using oscillometric finger cuffs in children and young adults. A comparison with arm cuffs during general anaesthesia.Anaesthesia. 1994 Oct;49(10):895-9. doi: 10.1111/j.1365-2044.1994.tb04270.x. Anaesthesia. 1994. PMID: 7802191 Clinical Trial.
-
Continuous noninvasive pulse wave analysis using finger cuff technologies for arterial blood pressure and cardiac output monitoring in perioperative and intensive care medicine: a systematic review and meta-analysis.Br J Anaesth. 2020 Jul;125(1):25-37. doi: 10.1016/j.bja.2020.03.013. Epub 2020 May 29. Br J Anaesth. 2020. PMID: 32475686
-
Fifteen years experience with finger arterial pressure monitoring: assessment of the technology.Cardiovasc Res. 1998 Jun;38(3):605-16. doi: 10.1016/s0008-6363(98)00067-4. Cardiovasc Res. 1998. PMID: 9747429 Review.
Cited by
-
Continuous mean arterial pressure measurement in the fingers: the influence of local arm cooling.Med Biol Eng Comput. 2001 Sep;39(5):584-9. doi: 10.1007/BF02345150. Med Biol Eng Comput. 2001. PMID: 11712656
References
MeSH terms
LinkOut - more resources
Full Text Sources