Comparative Study
. 1998 Sep 15;23(18):1946-55; discussion 1955-6.
doi: 10.1097/00007632-199809150-00005.

Biomechanical comparison of C1-C2 posterior fixations. Cable, graft, and screw combinations

Affiliations
Comparative Study

Biomechanical comparison of C1-C2 posterior fixations. Cable, graft, and screw combinations

S Naderi et al. Spine (Phila Pa 1976). .

Abstract

Study design: Four combinations of cable-graft-screw fixation at C1-C2 were compared biomechanically in vitro using nondestructive flexibility testing. Each specimen was instrumented successively using each fixation combination.

Objectives: To determine the relative amounts of movement at C1-C2 after instrumentation with various combinations of one or two transarticular screws and a posterior cable-secured graft. Also to determine the role of each component of the construct in resisting different types of loading.

Summary of background data: Spinal stiffness increases after instrumentation with two transarticular screws plus a posterior wire-graft compared with a wire-graft alone. Other C1-C2 cable-graft-screw combinations have not been tested.

Methods: Eight human cadaveric occiput-C3 specimens were loaded nondestructively with pure moments, and nonconstrained motion at C1-C2 was measured. The instrumented states tested were a C1-C2 interposition graft attached with multistranded cable; a cable-graft plus one transarticular screw; two transarticular screws alone; and a cable-graft plus two transarticular screws.

Results: The transarticular screws prevented lateral bending and axial rotation better than the posterior cable-graft. The cable-graft prevented flexion and extension better than the screws. Increasing the number of fixation points often significantly decreased the rotation and translation (paired t test; P < 0.05). Axes of rotation shifted from their normal location toward the hardware.

Conclusions: It is mechanically advantageous to include as many fixation points as possible when atlantoaxial instability is treated surgically.