[Safety considerations to avoid current-induced skin burns in MRI procedures]
- PMID: 9793133
- DOI: 10.1007/s001170050421
[Safety considerations to avoid current-induced skin burns in MRI procedures]
Abstract
The safety aspects of radiological methods continue to evolve. In this paper we report on two cases of skin burns in MRI caused by induced electrical current. A second- and a third-degree skin burn occurred during imaging in a 1.5 T system. The electromagnetic radiofrequency field inadvertently led to electrical currents caused by a conducting loop through the extremities and trunk. Skin burns induced by electrical current may occur in extremely rare cases even with standard MR imaging protocols operating within all current safety guidelines by inadvertently forming a closed conducting loop. By avoiding focal skin to skin contact of the extremities, this extremely rare adverse event can be avoided.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical