. 1998 Nov;209(2):381-5.
doi: 10.1148/radiology.209.2.9807562.

Internally cooled power laser for MR-guided interstitial laser-induced thermotherapy of liver lesions: initial clinical results

Affiliations

Internally cooled power laser for MR-guided interstitial laser-induced thermotherapy of liver lesions: initial clinical results

T J Vogl et al. Radiology. 1998 Nov.

Abstract

Purpose: To evaluate experimentally and clinically an internally cooled power laser system for percutaneous treatment of liver metastases, with magnetic resonance (MR) imaging guidance, to increase the volume of coagulative necrosis with single laser beam applications.

Materials and methods: The power laser system consisted of standard cannulation paraphernalia and a specially designed 7-F protective catheter for cooling the laser tip during treatment to avoid carbonization. A microdome laser beam applicator with a laser tip diameter of 0.95 mm kept the entire device as small as possible. After the unit was tested in 40 porcine livers, 532 laser beam applications (mean power, 27.5 W [range, 22.1-30.0 W]; mean application time, 19.8 minutes [range, 14-30 minutes]) were performed with MR guidance in 127 patients with 318 liver metastases. The safety of the procedure and the volume of induced necrosis were evaluated.

Results: In vitro, cuboid areas of coagulative necrosis with a mean greatest diameter of 4.3 cm +/- 0.5 (SD) were demonstrated after single applications. In vivo, MR thermometry allowed accurate demarcation of changes induced by heat, with a mean diameter of necrosis of 3.3 cm +/- 1.4. No carbonization or vaporization of tissue or damage of equipment occurred during treatment.

Conclusion: This power laser system can increase the volume of laser-induced necrosis in liver metastases with a single application, thereby simplifying and accelerating the treatment of larger lesions.