Liver lesions: improved detection with dual-detector-array CT and routine 2.5-mm thin collimation
- PMID: 9807568
- DOI: 10.1148/radiology.209.2.9807568
Liver lesions: improved detection with dual-detector-array CT and routine 2.5-mm thin collimation
Abstract
Purpose: To determine the feasibility and clinical benefit of routine performance of helical computed tomography (CT) with 2.5-mm collimation for the detection of liver lesions.
Materials and methods: Twenty patients with small (< or = 10-mm-diameter) liver lesions (total number of lesions, 167) were evaluated with a dual-detector-array CT scanner during the portal venous phase of contrast material enhancement. The acquisition was performed with 2.5-mm collimation during a single breath hold. The identical data set was used to perform reconstructions with 2.5-mm, 5.0-mm, 7.5-mm, and 10.0-mm section thicknesses with 50% section overlap. Each set of images was evaluated by three radiologists to determine lesion detection rates and conspicuity.
Results: Use of 2.5-mm-thick sections resulted in a 46% increase in detection rate versus use of 10.0-mm-thick sections (167 lesions vs 90 lesions), a 33% increase versus use of 7.5-mm-thick sections (167 vs 112), and an 18% increase versus use of 5-mm-thick sections (167 vs 137). Lesion conspicuity and radiologist confidence in lesion detection and characterization of lesion margins increased as section thickness decreased.
Conclusion: CT of the liver can be performed routinely with 2.5-mm collimation with a dual-detector CT system, yielding greater conspicuity of small lesions and improved lesion detection.