An evaluation of the Haema-Count MK-40 blood counting system
- PMID: 983996
- DOI: 10.1093/ajcp/66.5.883
An evaluation of the Haema-Count MK-40 blood counting system
Abstract
The Haema-Count MK-40, a semiautomated blood counting system, determines hemoglobin, erthrocyte count, leukocyte count, and hematocrit values of blood cell suspensions prepared with a small automatic pipetter-diluter. It is similar to the MK-3, which the authors previously evaluated, but has automatic coincidence correlation and a modified prime and rinse cycle. Its performance was compared with those of standard methods, i.e., the single-channel Coulter Counter and manual cyanmethemoglobin and microhematocrit methods. Precisions for hemoglobin determinations and for leukocyte and erythrocyte counts were equal to those of the reference (comparative) methods. Patients comparisons for those determinations had only small intermethod variability and small clinically insignificant biases. The hematocrit channel was the least precise. With a modified method of calibration, the authors obtained patient comparisons without statistically significant bias from the microhematocrit. Calibration was stable for all channels during the course of the study.
Similar articles
-
A report on experience with an automatic blood counting machine.J Clin Pathol. 1969 Jan;22(1):19-27. doi: 10.1136/jcp.22.1.19. J Clin Pathol. 1969. PMID: 5784690 Free PMC article.
-
Evaluation of the haema-count MK-3 blood counting system.Am J Clin Pathol. 1974 Jun;61(6):959-70. Am J Clin Pathol. 1974. PMID: 4841272 No abstract available.
-
[Evaluation of the Coulter Counter Model S].Rinsho Byori. 1973 Jun;21(6):514-20. Rinsho Byori. 1973. PMID: 4738352 Japanese. No abstract available.
-
Clinical implications of automation in cell counting systems.Clin Lab Haematol. 1979;1(1):1-12. doi: 10.1111/j.1365-2257.1979.tb00585.x. Clin Lab Haematol. 1979. PMID: 394902 Review. No abstract available.
-
Automated differential counting.Blood Rev. 1991 Jun;5(2):78-83. doi: 10.1016/0268-960x(91)90038-e. Blood Rev. 1991. PMID: 1717085 Review. No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources