Wear of congruent meniscal bearings in unicompartmental knee arthroplasty: a retrieval study of 16 specimens
- PMID: 9853488
- DOI: 10.1302/0301-620x.80b6.8176
Wear of congruent meniscal bearings in unicompartmental knee arthroplasty: a retrieval study of 16 specimens
Abstract
Many designs of unicompartmental knee replacement show early and mid-term failure due to polyethylene wear. We studied the wear rate of congruent polyethylene meniscal bearings retrieved from failed Oxford unicompartmental knee replacements. We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring their thickness and comparing it with that of 14 unused bearings. The mean rate of penetration, which included the effects of wear at both upper and lower surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5 mm wore no faster than thicker models, but ten with evidence of impingement had greater wear. The six bearings with no impingement showed a mean rate of penetration of 0.01 mm per year. In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear.
Similar articles
-
Polyethylene wear in meniscal knee replacement. A one to nine-year retrieval analysis of the Oxford knee.J Bone Joint Surg Br. 1992 Mar;74(2):228-32. doi: 10.1302/0301-620X.74B2.1544958. J Bone Joint Surg Br. 1992. PMID: 1544958
-
The Oxford medial unicompartmental arthroplasty: a ten-year survival study.J Bone Joint Surg Br. 1998 Nov;80(6):983-9. doi: 10.1302/0301-620x.80b6.8177. J Bone Joint Surg Br. 1998. PMID: 9853489
-
Polyethylene wear in Oxford unicompartmental knee replacement: a retrieval study of 47 bearings.J Bone Joint Surg Br. 2010 Mar;92(3):367-73. doi: 10.1302/0301-620X.92B3.22491. J Bone Joint Surg Br. 2010. PMID: 20190307
-
Failure mechanisms of unicompartmental knee replacement: the impact of changes in operative technique and component design.Semin Arthroplasty. 1991 Jan;2(1):23-8. Semin Arthroplasty. 1991. PMID: 10150059 Review.
-
Development and clinical application of meniscal unicompartmental arthroplasty.Proc Inst Mech Eng H. 2007 Jan;221(1):47-59. doi: 10.1243/09544119JEIM251. Proc Inst Mech Eng H. 2007. PMID: 17315768 Review.
Cited by
-
A structured evaluation of the symptomatic medial Oxford unicompartmental knee arthroplasty (UKA).EFORT Open Rev. 2021 Oct 19;6(10):850-860. doi: 10.1302/2058-5241.6.200105. eCollection 2021 Oct. EFORT Open Rev. 2021. PMID: 34760285 Free PMC article. Review.
-
[AMC unicondylar prosthesis].Orthopade. 2004 Nov;33(11):1267-76. doi: 10.1007/s00132-004-0727-z. Orthopade. 2004. PMID: 15480541 German.
-
Return to sport after total or unicompartmental knee arthroplasty: An informative guide for residents to patients.EFORT Open Rev. 2017 Dec 15;2(12):496-501. doi: 10.1302/2058-5241.2.170037. eCollection 2017 Dec. EFORT Open Rev. 2017. PMID: 29387472 Free PMC article.
-
Outcome of Unicompartmental Knee Arthroplasty: A Systematic Review of Comparative Studies between Fixed and Mobile Bearings Focusing on Complications.Knee Surg Relat Res. 2015 Sep;27(3):141-8. doi: 10.5792/ksrr.2015.27.3.141. Epub 2015 Sep 1. Knee Surg Relat Res. 2015. PMID: 26389066 Free PMC article. Review.
-
The Oxford unicompartmental knee prosthesis: an independent 10-year survival analysis.Knee Surg Sports Traumatol Arthrosc. 2006 Jan;14(1):40-5. doi: 10.1007/s00167-005-0621-1. Epub 2005 May 14. Knee Surg Sports Traumatol Arthrosc. 2006. PMID: 15895294
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical