Comparative Study
. 1998 Dec;20(6):685-90.
doi: 10.1097/00007691-199812000-00019.

Evaluation of an immunoassay (EMIT) for mycophenolic acid in plasma from renal transplant recipients compared with a high-performance liquid chromatography assay

Affiliations
Comparative Study

Evaluation of an immunoassay (EMIT) for mycophenolic acid in plasma from renal transplant recipients compared with a high-performance liquid chromatography assay

J L Beal et al. Ther Drug Monit. 1998 Dec.

Abstract

Mycophenolic acid is an immunosuppressant administered as a bioavailable ester, mycophenolate mofetil. The pharmacokinetics of mycophenolic acid have been reported to be variable. Accurate measurement of concentrations of this drug could be important to adjust doses. The aim of this study was to compare the enzyme-multiplied immunoassay technique (EMIT [Dade Behring; San Jose, CA, U.S.A.]) for mycophenolic acid with a high-performance liquid chromatographic (HPLC) assay using samples collected from renal transplant recipients. The HPLC assay used solid phase extraction and a C18 stationary phase with ultraviolet (UV) detection (254 nm). The immunoassay required no manual sample preparation. Plasma samples (n=102) from seven patients, collected at various times after a dose, were analyzed using both methods. Both assays fulfilled quality-control criteria. Higher concentrations were consistently measured in patient samples when using EMIT. The mean (+/-standard deviation [SD]) bias (EMIT-HPLC) was 1.88+/-0.86 mg/L. The differences in concentrations were higher in the middle of a dosage interval, suggesting that a metabolite might have been responsible for overestimation. Measurement of glucuronide concentrations by HPLC demonstrated only a weak correlation between assay differences and glucuronide concentrations. If the crossreacting substance is active, EMIT could provide a superior measure of immunosuppression; if inactive, further work is needed to improve antibody specificity. In conclusion, it was found that EMIT overestimates the concentration of mycophenolic acid in plasma samples from renal transplant recipients compared with HPLC analysis.