. 1998 Jul;10(7):569-72.
doi: 10.1097/00042737-199807000-00008.

Validation of [13C]urea breath test for Helicobacter pylori using a simple gas chromatograph-mass selective detector

Affiliations

Validation of [13C]urea breath test for Helicobacter pylori using a simple gas chromatograph-mass selective detector

H S Lee et al. Eur J Gastroenterol Hepatol. 1998 Jul.

Abstract

Objective: Isotope ratio mass spectrometry (IRMS) is the accepted method for accurately measuring the 13CO2:12CO2 ratio in the non-invasive and non-radioactive [13C]urea breath test (13C-UBT) for Helicobactor pylori. The IRMS instrument, an expensive and highly specialized analyser, is rarely available. The objective of this project was to modify and validate the use of a simple bench-top gas chromatograph-mass selective detector (GC-MSD) for 13C-UBT.

Methods: Breath samples from 71 patients were taken at baseline and 30 min after ingestion of 100 mg [13C]urea. The breath samples were analysed using GC-MSD in the selected ion monitoring mode. The reference 13CO2:12CO2 ratio was from NBS19 obtained from the US National Institute of Standards and Technology. 13CO2:12CO2 ratios of the breath samples were determined. Excess delta per thousand (per mil, delta/thousand) of the 30 min sample over the baseline (deltadelta/thousand) of > or = 6deltadelta/thousand was considered H. pylori positive. Results from 13C-UBT and histology determined blind to each other were compared.

Results: The coefficient of variation of the reference 13CO2:12CO2 ratio was 0.06%. Using histology as the 'gold standard', the sensitivity (97.9%) and specificity (95.8%) of the GC-MSD 13C-UBT were comparable to those of other methods of H. pylori diagnosis.

Conclusion: A gas chromatograph coupled to a mass selective detector that is available in many analytical and biomedical laboratories can be used for the 13C-UBT. This method will increase the availability and reduce the cost of this non-invasive, non-radioactive diagnostic test.