Stimulus duration, neural adaptation, and sweep visual evoked potential acuity estimates
- PMID: 9856787
Stimulus duration, neural adaptation, and sweep visual evoked potential acuity estimates
Abstract
Purpose: Results in several studies have suggested that the visual evoked potential (VEP) amplitude can vary with stimulus duration. The purpose of this study was to determine whether acuity estimates obtained by extrapolation of the sweep VEP are altered by this adaptation effect.
Methods: Sweep VEP data were obtained from 16 healthy observers under binocular viewing conditions. Data were acquired with a commercially available VEP unit using standard electrode recording techniques. Three sweeps (high spatial frequencies, medium spatial frequencies, and low spatial frequencies) were run. The subjects' visual acuity at the monitor distance was 6/6 for the high spatial frequency sweep. For the medium and low spatial frequency sweeps, the subjects were dioptrically blurred to 6/15 (medium spatial frequencies) or 6/30 (low spatial frequencies) at the monitor distance. Each sweep consisted of six spatial frequencies (contrast 80%; temporal frequency (TF) = 7.5 Hz; screen luminance = 100 candela [cd]/m2). For each spatial frequency, the stimulus duration was 8 seconds, partitioned into 1-second bins. A minimum of eight sweeps were obtained per subject. An acuity estimate was obtained for each second's data by fitting a line to the high spatial frequencies (excluding noise) and extrapolating this line to the x-axis. With this technique, estimates could not be obtained for 29 of 384 possible acuities.
Results: The sweep VEP acuities for the 16 subjects did not change significantly over the 8 seconds of data collection for the high, medium, or low spatial frequency sweep (repeated measures analysis of variance [ANOVA]: high, P = 0.25; medium, P = 0.50; low, P = 0.23). In any given subject, there was a 1- to 2-octave range in acuity estimates over the 8 seconds of stimulus presentation (high, 1.23+/-0.417 octaves; medium, 1.41+/-0.593 octaves; low, 1.52+/-0.475 octaves; mean +/- SD).
Conclusions: These results suggest that there is not a significant change in sweep VEP acuity estimates over an 8-second stimulus presentation. Thus, neural adaptation does not significantly affect the clinical use of the sweep VEP.
Similar articles
-
Methods of visual acuity determination with the spatial frequency sweep visual evoked potential.Doc Ophthalmol. 2004 Nov;109(3):239-47. doi: 10.1007/s10633-004-8053-7. Doc Ophthalmol. 2004. PMID: 15957609
-
Sweep visual evoked potential grating acuity thresholds paradoxically improve in low-luminance conditions in children with cortical visual impairment.Invest Ophthalmol Vis Sci. 2006 Jul;47(7):3220-4. doi: 10.1167/iovs.05-1252. Invest Ophthalmol Vis Sci. 2006. PMID: 16799070
-
[Measuring vernier acuity and motion perception with the parameter sweep VEP].Klin Monbl Augenheilkd. 1996 Jan;208(1):11-7. doi: 10.1055/s-2008-1035161. Klin Monbl Augenheilkd. 1996. PMID: 8839339 German.
-
The technique, validity and clinical use of the sweep VEP.Ophthalmic Physiol Opt. 2008 Sep;28(5):393-403. doi: 10.1111/j.1475-1313.2008.00591.x. Ophthalmic Physiol Opt. 2008. PMID: 18761477 Review.
-
[Usefulness of visual evoked potentials in visual acuity examination].Klin Oczna. 2001;103(2-3):141-5. Klin Oczna. 2001. PMID: 11873414 Review. Polish.
Cited by
-
Threshold determination in sweep VEP and the effects of criterion.Doc Ophthalmol. 2009 Oct;119(2):109-21. doi: 10.1007/s10633-009-9177-6. Epub 2009 Jun 24. Doc Ophthalmol. 2009. PMID: 19554357
-
Comparing enfant and PowerDiva sweep visual evoked potential (sVEP) acuity estimates.Doc Ophthalmol. 2014 Oct;129(2):105-14. doi: 10.1007/s10633-014-9457-7. Epub 2014 Aug 24. Doc Ophthalmol. 2014. PMID: 25150841
-
VEP estimation of visual acuity: a systematic review.Doc Ophthalmol. 2021 Feb;142(1):25-74. doi: 10.1007/s10633-020-09770-3. Epub 2020 Jun 2. Doc Ophthalmol. 2021. PMID: 32488810 Free PMC article.
-
Methods of visual acuity determination with the spatial frequency sweep visual evoked potential.Doc Ophthalmol. 2004 Nov;109(3):239-47. doi: 10.1007/s10633-004-8053-7. Doc Ophthalmol. 2004. PMID: 15957609
-
Reliability of acuities determined with the sweep visual evoked potential (sVEP).Doc Ophthalmol. 2012 Apr;124(2):99-107. doi: 10.1007/s10633-012-9312-7. Epub 2012 Jan 20. Doc Ophthalmol. 2012. PMID: 22262233
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous