Experimental and clinical evaluation of incision size and shape following forceps and injector implantation of a three-piece high-refractive-index silicone intraocular lens
- PMID: 9865623
- DOI: 10.1007/s004170050181
Experimental and clinical evaluation of incision size and shape following forceps and injector implantation of a three-piece high-refractive-index silicone intraocular lens
Abstract
Background: Previous studies have demonstrated that incision sizes required for insertion of various foldable intraocular lenses (IOLs) vary according to the IOL/inserter combination. The purpose of this study was to compare incision sizes and wound shape for both forceps and injector implantation of high-refractive-index silicone IOLs.
Methods: In fresh human cadaver eyes, limbal corneal tunnel incisions were created, and 12 foldable high-refractive-index silicone (dimethyldiphenylsiloxane) IOLs were inserted in a randomized fashion using either a forceps or an injector. Using incision calipers, internal and external measurements of the tunnel incisions were obtained before and after IOL insertion. Scanning electron microscopy (SEM) was performed on selected corneas following IOL insertion. Additionally, in 12 cataract procedures, the incision sizes following forceps or injector implantation were evaluated intraoperatively.
Results: In the experimental setting, the external and internal tunnel widths (in mm) before insertion were 3.05 mm (+/-0.07) and 3.02 mm (+/-0.03), respectively, with the forceps and 3.06 mm (+/-0.04) and 3.01 mm (+/-0.04) with the injector. Following IOL implantation, the external and internal incision sizes were 3.33 mm (+/-0.07) and 3.33 mm (+/-0.04) with the forceps and 3.32 mm (+/-0.08) and 3.33 mm (+/-0.07) with the injector. SEM showed tearing of corneal structures after implantation through the smallest possible incisions with both devices. In the clinical study, the incision sizes before and after implantation were 3.23 mm (+/-0.10) and 3.36 mm (+/-0.06) with the forceps and 3.11 mm (+/-0.08) and 3.21 mm (+/-0.10) with the injector.
Conclusions: This study confirms that high-refractive-index silicone IOLs provide one of the smallest currently achievable incisions before and after IOL implantation. In clinical use, the new injector reduces the mean incision size required by approximately 0.1 mm.
Similar articles
-
Incision sizes with 5.5 mm total optic, 3-piece foldable intraocular lenses.J Cataract Refract Surg. 2000 Dec;26(12):1765-72. doi: 10.1016/s0886-3350(00)00526-5. J Cataract Refract Surg. 2000. PMID: 11134877 Clinical Trial.
-
Incision sizes for foldable intraocular lenses.Ophthalmology. 1997 Aug;104(8):1277-86. doi: 10.1016/s0161-6420(97)30147-x. Ophthalmology. 1997. PMID: 9261314
-
Incision width after phacoemulsification with foldable intraocular lens implantation.J Cataract Refract Surg. 2000 Feb;26(2):237-41. doi: 10.1016/s0886-3350(99)00346-6. J Cataract Refract Surg. 2000. PMID: 10683791
-
Incision sizes before and after implantation of 6-mm optic foldable intraocular lenses using Monarch and Unfolder injector systems.Ophthalmology. 2005 Jan;112(1):58-66. doi: 10.1016/j.ophtha.2004.06.030. Ophthalmology. 2005. PMID: 15629821 Review.
-
[Sutureless scleral intraocular lens fixation: report of nine cases and literature review].J Fr Ophtalmol. 2013 Oct;36(8):658-68. doi: 10.1016/j.jfo.2012.09.009. Epub 2013 Jul 25. J Fr Ophtalmol. 2013. PMID: 23891322 Review. French.
Cited by
-
Alpha-Glycerylphosphorylcholine and D-Panthenol Eye Drops in Patients Undergoing Cataract Surgery.J Ophthalmol. 2022 Jun 7;2022:1951014. doi: 10.1155/2022/1951014. eCollection 2022. J Ophthalmol. 2022. PMID: 35711284 Free PMC article.
-
[Incisions for biaxial and coaxial microincision cataract surgery].Ophthalmologe. 2010 Feb;107(2):108-15. doi: 10.1007/s00347-009-1980-7. Ophthalmologe. 2010. PMID: 20107810 Review. German.
-
Multicenter Evaluation of Time, Operational, and Economic Efficiencies of a New Preloaded Intraocular Lens Delivery System versus Manual Intraocular Lens Delivery.Clin Ophthalmol. 2021 Feb 16;15:591-599. doi: 10.2147/OPTH.S263658. eCollection 2021. Clin Ophthalmol. 2021. PMID: 33623360 Free PMC article.
-
Evaluation of Preloaded IOL Delivery System and Hydrophobic Acrylic Intraocular Lens in Cataract Surgery.Open Ophthalmol J. 2018 Jun 14;12:94-103. doi: 10.2174/1874364101812010094. eCollection 2018. Open Ophthalmol J. 2018. PMID: 30008971 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources