Clinical Trial
. 1998 Dec;116(12):1579-82.
doi: 10.1001/archopht.116.12.1579.

Quantitative comparison of posterior capsule opacification after polymethylmethacrylate, silicone, and soft acrylic intraocular lens implantation

Affiliations
Clinical Trial

Quantitative comparison of posterior capsule opacification after polymethylmethacrylate, silicone, and soft acrylic intraocular lens implantation

H Hayashi et al. Arch Ophthalmol. 1998 Dec.

Abstract

Objective: To quantitatively compare the extent of posterior capsule opacification (PCO) after polymethylmethacrylate (PMMA), silicone, and soft acrylic intraocular lens implantation.

Patients and methods: A total of 240 eyes from 240 patients undergoing implant surgery were randomized into 3 groups based on the type of lens implanted: PMMA, silicone, and soft acrylic. The density value of PCO in 185 eyes was quantitated approximately 2 years after surgery by a new measurement method using the Scheimpflug videophotography system.

Results: Twenty-one eyes (30.4%) in the PMMA group, 4 (5.7%) in the silicone group, and 2 (2.7%) in the acrylic group had already undergone Nd:YAG laser posterior capsulotomy. The mean +/- SD PCO values were 26.3 +/- 12.2 computer-compatible tape steps (CCT) in the PMMA group, 12.0 +/- 8.3 CCT in the silicone group, and 16.0 +/- 10.3 CCT in the acrylic group. The PCO value in the PMMA group was significantly greater than that in the silicone or acrylic group (P < .001). The visual acuity loss in the PMMA group was also greater than that in the silicone or acrylic group (P < .001).

Conclusion: Based on the PCO value and capsulotomy rate, the PCO was more extensive with the PMMA lens than with either the silicone or soft acrylic lens, which led to visual acuity loss.