Comparison of NucliSens and Roche Monitor assays for quantitation of levels of human immunodeficiency virus type 1 RNA in plasma
- PMID: 9889240
- PMCID: PMC84337
- DOI: 10.1128/JCM.37.2.447-449.1999
Comparison of NucliSens and Roche Monitor assays for quantitation of levels of human immunodeficiency virus type 1 RNA in plasma
Abstract
We compared the performance of Organon Teknika's NucliSens and Roche Diagnostic Systems' Monitor quantitative human immunodeficiency type 1 RNA assays. Both had similar linearity and sensitivity over most of the dynamic range of the assays, although the Monitor assay was superior at the low range of RNA values while the NucliSens assay was more consistent at higher RNA values. NucliSens generally showed less interassay variability.
Figures

References
-
- Brambilla D, Leung S, Lew J, Todd J, Herman S, Cronin M, Shapiro D E, Bremer J, Hanson C, Hillyer G V, Sherry G D M, Sperling R S, Coombs R W, Reichelderfer P S. Absolute copy number and relative change in determinations of human immunodeficiency virus type 1 RNA in plasma: effect of an external standard on kit comparisons. J Clin Microbiol. 1998;36:311–314. - PMC - PubMed
-
- Dyer J R, Pilcher C D, Shepard R, Schock J, Eron J J, Fiscus S A. Abstr. 42162. 12th World AIDS Conference, Geneva, Switzerland. 1998.
-
- Finzi D, Siliciano R F. Viral dynamics in HIV-1 infection. Cell. 1998;93:665–671. - PubMed
-
- Griffith B P, Rigsby M O, Garner R B, Gordon M M, Chacko T M. Comparison of the Amplicor HIV-1 Monitor test and the nucleic acid sequence-based amplification assay for quantitation of human immunodeficiency virus RNA in plasma, serum, and plasma subjected to freeze-thaw cycles. J Clin Microbiol. 1997;35:3288–3291. - PMC - PubMed
Publication types