. 1976;2(4):242-56.
doi: 10.1159/000119621.

Hydrodynamics of shunt valves

Hydrodynamics of shunt valves

H D Portnoy et al. Childs Brain. 1976.

Abstract

A study of differential pressure valves (DPV) used in the treatment of hydrocephalus was undertaken to determine their pressure-flow characteristics and compatability with the antisiphon valve (ASV). DPV could be classified into two groups: low resistance valves (LRV) and high resistance valves (HRV). The LRV maintains intraventricular pressure (IVP) near the closing pressure (CP) of the valve by permitting a high flow whenever CP is exceeded. The HRV regulates IVP by attempting to match inflow with some point on the pressure-flow curve of the valve. These characteristics were lost unless valve outlet pressure was maintained at atmospheric pressure. This could be accomplished by using a proximal DPV with an ASV at the DPV outlet, thus converting the DPV into a gauge pressure valve and preventing the 'siphon effect' seen with the use of a DPV alone.