The Otto Aufranc Award. Image guided navigation system to measure intraoperatively acetabular implant alignment

Affiliations
Clinical Trial

The Otto Aufranc Award. Image guided navigation system to measure intraoperatively acetabular implant alignment

A M DiGioia et al. Clin Orthop Relat Res. 1998 Oct.

Abstract

There has been little clinical research to examine the effects of patient positioning and pelvic motion on the alignment of the acetabular implant during total hip replacement surgery. Until now, no tools were capable of accurately measuring these variables during the actual procedure. As part of a broader program in medical robotics and computer assisted surgery, a clinical system has been developed that includes several enabling technologies. The hip navigation system (HipNav) continuously and precisely measures pelvic location and tracks relative implant alignment intraoperatively. HipNav technology is used to gauge current clinical practice and provide intraoperative feedback to surgeons with the goal of improving the precision and accuracy of acetabular alignment during total hip replacement. This system provides surgeons with a new class of image guided measurement tools and assist devices. These tools successfully were introduced into the clinical practice of surgery with results showing the following: (1) There exist unpredictable and large variations in the initial position of patients' pelves on the operating room table and significant pelvic movement during surgery and during intraoperative range of motion testing; (2) current mechanical acetabular alignment guides do not account for these variations, and result in variable and in the majority of cases unacceptable acetabular alignment; and (3) press fitting oversized acetabular components influences the final cup orientation.