Single-point mutations causing more than 100-fold underestimation of human immunodeficiency virus type 1 (HIV-1) load with the Cobas TaqMan HIV-1 real-time PCR assay
- PMID: 19193839
- PMCID: PMC2668364
- DOI: 10.1128/JCM.02204-08
Single-point mutations causing more than 100-fold underestimation of human immunodeficiency virus type 1 (HIV-1) load with the Cobas TaqMan HIV-1 real-time PCR assay
Abstract
Systematic sequence analysis of human immunodeficiency virus type 1 (HIV-1) variants with RNA levels underestimated by the Cobas TaqMan HIV-1 assay demonstrated that mutations at a single position of the downstream primer can lead to the underestimation of HIV-1 RNA levels by more than 2 log and to false-negative results in minipool screening of blood donors. Mutations at this position are found in about 2% of all HIV-1 M gag sequences.
Figures
References
-
- Colson, P., C. Solas, J. Moreau, A. Motte, M. Henry, and C. Tamalet. 2007. Impaired quantification of plasma HIV-1 RNA with a commercialized real-time PCR assay in a couple of HIV-1-infected individuals. J. Clin. Virol. 39226-229. - PubMed
-
- Damond, F., B. Roquebert, A. Bénard, G. Collin, M. Miceli, P. Yéni, F. Brun-Vezinet, and D. Descamps. 2007. Human immunodeficiency virus type 1 (HIV-1) plasma load discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 assays. J. Clin. Microbiol. 453436-3438. - PMC - PubMed