. 2026 Feb 16:S0735-1097(26)05423-9.
doi: 10.1016/j.jacc.2026.02.5063. Online ahead of print.

Six-Year Outcomes After Transcatheter vs Surgical Aortic Valve Replacement in Low-Risk Patients With Aortic Stenosis

Collaborators, Affiliations
Free article

Six-Year Outcomes After Transcatheter vs Surgical Aortic Valve Replacement in Low-Risk Patients With Aortic Stenosis

John K Forrest et al. J Am Coll Cardiol. .
Free article

Abstract

Background: The Evolut Low Risk trial enrolled patients with severe aortic stenosis at low surgical risk. Annual follow-up is planned for 10 years, evaluating the composite of all-cause mortality or disabling stroke and key secondary endpoints.

Objectives: Our prespecified objective was to report the 6-year clinical outcomes of transcatheter aortic valve replacement (TAVR) vs surgery from the Evolut Low Risk trial. Given an increase in reintervention rates at 6 years, we performed additional analyses in available 7-year data.

Methods: Low-risk patients with severe symptomatic aortic stenosis were randomized to TAVR or surgery from 2016-2019. Prespecified analyses at 6 years included annual follow-up of clinical outcomes reported as Kaplan-Meier estimates with log-rank test. Because the trial enrolled patients over several years, at the time of data lock, a majority of patients had completed 7-year follow-up. Given an increased reintervention rate at 6 years in the TAVR arm, we performed additional analysis of 7-year data available at the time of the database lock. Reintervention rates are reported as cumulative incidence.

Results: A total of 1,414 patients underwent an attempted implantation (730 TAVR, 684 surgery). At 6 years, the composite endpoint of all-cause mortality or disabling stroke was 23.3% for TAVR and 20.4% for surgery (difference: 2.8% [95% CI: -1.9% to 7.6%]; P = 0.43). All-cause mortality with vital status sweep at 6 years was 23.3% (95% CI: 20.6%-26.4%) for TAVR and 20.2% (95% CI: 17.4%-23.3%) for surgery (P = 0.24). The reintervention rate at 6 years was 5.5% for TAVR and 3.3% for surgery (sHR: 1.66 [95% CI: 0.96-2.86]; P = 0.07). Using available 7-year follow-up (555 TAVR and 480 surgery), the reintervention rate for TAVR was 9.8% and for surgery was 6.0% (sHR: 1.68 [95% CI: 1.10-2.58]; P = 0.02). In the TAVR and surgery groups, the rate of reintervention for regurgitation was 5.6% vs 1.6% (sHR: 3.39 [95% CI: 1.62-7.07]; P < 0.001) and the rate of reintervention for stenosis was 3.6% vs 3.5% (sHR: 1.14 [95% CI: 0.61-2.15]; P = 0.70).

Conclusions: The 6-year results from the Evolut Low Risk trial show no significant difference in the composite endpoint of all-cause mortality or disabling stroke. At 6 and 7 years, the TAVR arm had a higher reintervention rate compared with surgery, driven by an increased incidence of aortic regurgitation. (Medtronic Evolut Transcatheter Aortic Valve Replacement in Low Risk Patients; NCT02701283).

Keywords: TAVR; aortic stenosis; low-risk; supra-annular self-expanding; surgery.

PubMed Disclaimer

Conflict of interest statement

Funding Support and Author Disclosures This study was funded by Medtronic (Medtronic Evolut Transcatheter Aortic Valve Replacement in Low Risk Patients). Dr Forrest has received grant support/research contracts and consultant fees/honoraria/speakers bureau fees from Edwards Lifesciences and Medtronic. Dr Yakubov receives grants and personal fees from Medtronic and Boston Scientific. Dr Deeb serves on an advisory board for Medtronic; has received institutional grant support from Boston Scientific, Edwards Lifesciences, and Medtronic; and has received fees as a proctor for the Medtronic-sponsored SMART Trial. Dr Reardon has received research grants from Abbott, Boston Scientific, WL Gore Medical, and Medtronic.